pH/ISE and EC/TDS Benchtop Meter - IC-HI5521-02

ORDER ME IN
Product code: IC-HI5522-02

Backorder Now, once ordered your item will dispatch before the 12th of June 2025, call 1300 737 871 (NZ 0800 737 871 to confirm).

The IC-HI5522 is an advanced research grade, two-channel benchtop pH/ORP/ISE and EC/TDS/Salinity/Resistivity meter. The HI5522 is a two-channel meter that allows for simultaneous measure of pH, ORP, or ISE on one channel and EC, TDS, Salinity, or Resistivity on the other.
Regular price $2,400.76 AUD + GST
$2,640.84 AUD Inc. GST

Features:

Highly Customizable User Interface - The user interface of the HI5522 allows the user to show measurements in various modes: basic measurement with or without GLP information, real-time graphing, and logging data.

Capacitive Touch - The HI5522 features sensitive capacitive touch buttons for accurate keystrokes when navigating menus and screens. There are four dedicated keys that are used for routine operations including calibration and switching measurement modes and four virtual keys that change based upon use.

Choice of Calibration - Automatic buffer recognition, semiautomatic, and direct manual entry pH calibration options are available for calibrating up to five points, from a selection of eight standard buffers and up to five custom buffers.

GLP Data - The IC-HI5522 includes a GLP Feature that allows users to view calibration data and calibration expiration information at the touch of a key. Calibration data include date, time, buffers /standards used for calibration and slope characteristics. The offset is also displayed for pH electrodes.

CAL Check- CAL Check alerts users to potential problems during the calibration of the pH electrode. Indicators include -Electrode Dirty/Broken,-Buffer Contaminated-, electrode response time and the overall probe condition as a percentage that is based on the offset and slope characteristics.

Data Logging - Three selectable logging modes are available on the HI5522: automatic, manual, and AutoHold logging. Automatic and manual logs up to 100 lots with 50,000 records max/lot, with up to 100,000 total data points. Automatic logging features the option to save data according to sampling period and interval.


Specifications

pH:

pH Range: -2.000 to 20.000 pH
pH Resolution: 0.1, 0.01, 0.001 pH
pH Accuracy (@25C/77F): 0.1 pH, 0.01 pH, 0.002 pH 1 LSD
pH Calibration: automatic, up to five point calibration, eight standard buffers available (1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01,12.45) and five custom buffers
pH Temperature Compensation: automatic or manual from -20.0 to 120.0 C

mV:

mV Range: 2000 mV
mV Resolution: 0.1 mV
mV Accuracy: 0.2 mV 1 LSD
Relative mV Offset Range: 2000 mV

EC:

EC Range: 0.000 to 9.999 S/cm, 10.00 to 99.99 S/cm, 100.0 to 999.9 S/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC*
EC Resolution: 0.001 S/cm, 0.01 S/cm, 0.1 S/cm, 1 S/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy (@25C/77F): 1% of reading (0.01 S/cm)
EC Calibration: automatic standard recognition (0.000 S/cm, 84.00 S/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration

TDS:

TDS Range: 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor)
TDS Resolution: 0.001 ppm, 0.01 ppm, 0.1 ppm, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt
TDS Accuracy (@25C/77F): 1% of reading (0.01 ppm)
Resistivity Range: 1.0 to 99.9 O?cm; 100 to 999 O?cm; 1.00 to 9.99 KO?cm; 10.0 to 99.9 KO?cm; 100 to 999 KO?cm; 1.00 to 9.99 MO?cm; 10.0 to 100.0 MO?cm
Resistivity Resolution: 0.1 O?cm; 1 O?cm; 0.01 KO?cm; 0.1 KO?cm; 1 KO?cm; 0.01 MO?cm; 0.1 MO?cm*
Resistivity Accuracy: 2% of reading (1 O?cm)

Salinity:

Salinity Range: practical scale: 0.00 to 42.00 psu; natural seawater scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0%
Salinity Resolution: 0.01 for practical scale/natural seawater scale; 0.1% for percent scale
Salinity Accuracy (@25C/77F): 1% of reading
Salinity Calibration: percent scale?one-point (with HI7037 standard)

Temperature:

Temperature Range: -20.0 to 120.0 C, -4.0 to 248.0 F, 253.15 to 393.15 K
Temperature Resolution: 0.1 C, 0.1 F, 0.1 K
Temperature Accuracy: 0.2 C, 0.4 F, 0.2 K
Temperature Compensation: disabled, linear and non-linear (natural water)
Electrode/Probe: HI1131B glass body pH electrode with BNC connector and 1 m (3.3?) cable (included)
Temperature Probe: HI7662-T stainless steel temperature probe with 1 m (3.3?) cable (included)
GLP: calibration data including date, time, buffers used, offset and slope

ISE:

ISE Range: 1 x 10?6 to 9.99 x 10? concentration
ISE Resolution: 1; 0.1; 0.01; 0.001 concentration
ISE Accuracy @ 25 C/77 F: 0.5% (monovalent ions); 1% (divalent ions)
ISE Calibration Points: automatic, up to five point calibration, five fixed standard solutions available (0.1, 1, 10, 100, 1000 choice of concentration) and five user defined units


Our Reviews



Recently Viewed

Instrument Choice will respond in 3 business hours