The MA9071 electrolyte fill solution is designed for Milwaukee's polarographic dissolved oxygen probe. A high quality electrolyte solution is crucial to the performance of your dissolved oxygen probe.

“OPEN AIR” CALIBRATION PROCEDURE FOR THE MW600 DISSOLVED OXYGEN METER

1. Remove the membrane

Take the membrane off the end of the probe and dry the diode off.

2. Zero out the display

Connect the probe to the base unit and turn the unit on. Your display should read 0.0. If it does not, go to the position “0” screw on the front of the unit with the screw driver provided and through the access hole make the unit display read 0.0.

3. Fill the membrane

Go to the membrane and put the “O” ring inside the membrane on the shoulder at the base of the threads. Fill the membrane to just short of the “O” ring with MA9071 electrolyte solution. Be sure that there are no bubbles inside the membrane. (Air bubbles are the main cause of calibration failure)

4. Attach the membrane

Screw the membrane on to the probe.

5. Let stabilise

Turn the meter on and hang the probe off the end of a table pointing down and let the unit display stabilize (this may take 5 to 15 minutes)

6. Adjust

Push down and hold the “Cal” button on the front face of the meter and with the screwdriver go to the front Cal position “100%” and make the display read 100.


Our Reviews



Recently Viewed

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au