EcoScan DO 6 Dissolved Oxygen Meter with 3m cable electrode , 2 assembled membrane housing, 1 refill - EC-DO6-02K

ORDER ME IN
Product code: EC-DO6-02K

Backorder Now, once ordered your item will dispatch before the 22nd of May 2025, call 1300 737 871 (NZ 0800 737 871 to confirm).

EcoScan DO 6 Dissolved Oxygen Meter with 3m cable electrode , 2 assembled membrane cap housing, 1 refilling electrolyte & carrying kit set.
Regular price $1,999.50 AUD + GST
$2,199.45 AUD Inc. GST

Product Features

* User Customisation for user's flexibility like toggling of variousmodes from ppm, mg/l to % saturation, view electrode diagnosis, etc.
* Easy Calibration can be performed at 100% and/or 0% and mg/l or ppm against a known solution
* Automatic Temperature Compensation (ATC) guarantees the highest accuracy under varying temperature condition
* Galvanic Cell allows instantaneous measurements, producing quickresponses and eliminates polarisation; requires minimal maintenance]
* Automatically Compensates Salinity and Barometric Pressure after manual input for increased accuracy
* Hold Function momentarily freezes reading for easy viewing
* Auto-Power Off saves battery power after non-use
* Self-Diagnostic with message codes for easy troubleshooting
* Large Custom LCD provides optimum viewing even at a distance
* Protective Rubber Boot shields meter from accidental knocks and features a sturdy built-in stand for easy bench-top operation
* Rugged Carrying Case (Optional) provides handy on-site meter calibration and measurement

Applications

Aquaculture: Use to monitor oxygenlevels in catfish and shrimp farming; game stocking ponds; ornamentalfish tanks and ponds; and in other fish farming applications.

Industrial: Ideal for checks thequality of plant water intake and discharge, wastewater and watertreatment, recirculating systems and industrial process systems.

Environmental: Use to test water quality, monitor health of aquatic ecosystems, survey surface and ground water drinking supplies.

Educational: Ideal for quick, accurate DO readings in laboratories and schools.

 

Specifications

Product Specification Description
Dissolved Oxygen Range 0 to 20.00 mg/l or ppm
Resolution & Accuracy 0.01 mg/l or 0.01 ppm & 1.5% Full Scale
% Saturation of Oxygen 0.0 to 200/0 %
Resolution & Accuracy 0.1% & 1.5% Full Scale
Temperature Range 0.0 to 100.0 C
Resolution & Accuracy 0.1 C & 0.3 C
Salinity Correction 0.0 to 50.0 ppt
Barometric Pressure Correction 555 to 1499 mm Hg or 66.6 to 199.9 kPA
Method Automatic correction after manual input
No. of Calibration Points Up to 2 points; 100% in air and/or
0% in known solution
Special Functions Self-diagnostic; Electrode characteristics;
Hold & Auto-Power Off after20 minutes
Power Requirements 4 x 1.4V batteries; > 700 hrs

Our Reviews



Recently Viewed

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au