The HI11310 is a glass body refillable double junction pH electrode with a built-in temperature sensor for temperature compensated measurements in a single probe design. This probe connects to the meter with a 3.5 mm connector. The 3.5 mm connector is proprietary to meters that use digital electrodes.
The HI11310 electrode features a built-in microchip that stores sensor type, serial number, and calibration information. It also features a single ceramic
junction and the spherical pH sensing portion is made with high temperature glass. This design consideration is ideal for general purpose laboratory
use.
Features:
- Built-In Microchip: The built-in microchip stores sensor type, serial number and calibration information including date, time, offset, slope, probe condition
and buffers used. This information is automatically retrieved by edge once the electrode is plugged in. The ability to transfer information allows
for hot swapping of probes without having to recalibrate. All pH measurements are performed within the electrode and transferred digitally to the meter.
This overcomes any noise issues associated with the traditional high impedance analog measuring system. Electrical noise can be generated from a built
in temperature sensor and while working in a humid environment.
- Spherical Glass Tip: The spherical shaped tip design allows for a wide area of contact with the measured sample. This permits a faster electrode response
with a higher degree of stability. The HI10480 utilizes a low temperature (LT) glass formulation that has a resistance of around 50 mega ohms at 25
C. As the temperature of the glass decreases in the sample, the resistance of the LT glass will approach that of GP glass. If using GP glass, the resistance
would increase above the optimum range, resulting in increased impedance and ultimately affecting the measurement. The HI10480 is suitable to use with
samples that measure from 0 to 80 C.
- Double Junction Reference: A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte
of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver free. The double junction
design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the
sample are undesirable or silver precipitates
on the junction are likely to form.
- Glass Electrode Body: The glass electrode body is suitable for a wide range of applications due to its chemical resistance. The glass electrode is compatible
with many non-aqueous solvents and other aggressive chemicals. Glass is also resistant to many forms of radiation, such as ultraviolet radiation.
- Built-In Temperature Sensor: HI11310 features a built-in thermistor temperature sensor that is in the tip of the indicating pH electrode. A thermistor
temperature sensor provides for a high accuracy temperature reading and should be as close as possible to the indicating pH electrode in order to compensate
for the effect that temperature has on the membrane potential. By having an accurate reading it is possible to provide an accurate temperature compensated
reading.
Specifications
- Measurement Range pH: 0 to 13
- Reference Cell Type: Double
- Junction Type/Flow Rate: ceramic, single / 15-20 uL/h
- Refill Electrolyte: 3.5M KCl
- Body Material: Glass
- Tip / Shape: Spheric
- Matching Pin: No
- Max. Pressure: 0.1 bar
- Temperature Operating Range: -5 to 100 C (23 to 212 F)
- Body Length/Overall Length: 120 mm/175.5 mm
- Temperature Sensor: yes
- Outer Diameter: 12 mm
- Cable: Coaxial; 1 m (3.3 inch)
- Connector Type: 3.5 mm connector
- Amplifier: No
- Recommended Use: laboratory general purpose