Dual Channel GFX Vaccine Monitoring Kit with Graphic Screen and Audible Alarm - EL-GFX-VAC2

ORDER ME IN
Product code: IC-EL-GFX-VAC2

Backorder Now, once ordered your item will dispatch before the 23rd of May 2025, call 1300 737 871 (NZ 0800 737 871 to confirm).

TheEL-GFX-VAC2 is a thermistor probe data logger paired with glycol bottle probes for applications that require highly accurate temperature monitoring such as vaccination temperature monitoring systems.
Regular price $594.00 AUD + GST
$653.40 AUD Inc. GST
The EL-GFX-DTP data logger measures and stores up to 250,000 temperature readings from the supplied thermistor probes which are attached viathe sockets at the base of the unit. The user can easily set up the logging rate and start-time, and download the stored data by plugging the data loggerinto a PC's USB port and running the purpose designed software under Windows XP, Vista, 7 & 8.

These units come with external thermistor glycol bottles - allowing for discrete and continuous temperature monitoring of vaccines. EL-GFX-DTP-PROBE-G also includes a traceable calibration certificate.

The data logger features a dot-matrix LCD and three face-buttons to navigate through an on-screen menu. This menu provides the user with access to real-time trend analysis, data summaries and the ability to start, stop and restart the data logger without the need to connect the data logger to the host-PC.

Set up and data download of our EasyLog data loggers is enabled by our EL-WIN-USB software. compatible with 32-bit and 64-bit versions of Windows 7, 8 and 10


Features:

* 2 glycol filled bottles with thermistor probe on 3.0m cables
* Continuous temperature measurement with logging rates between 2 sec and 1 hr
* Micro USB interface for set-up and data download
* Probe measurement range of -40 to +100C (-40 to +212F)
* Accuracy of +/-0.1C (-10 to 70C), +/-0.3C (full range)
* Includes Calibration Certificate

Specifications

:

* Supplied probe measurement range: -40 (-40) to +100 (+212) C (F)
* Operating temperature range: -10 (-14) to +40 (+104) C (F)
* Internal resolution: 0.1C (0.2) (F)
* Accuracy (logger error): +/-0.1 (+/-0.2) C (F)
* Logging rate: Every 2s/Every 1hr

Multimedia:


Our Reviews



Recently Viewed

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au