How to Calibrate the IC860031 Benchtop Digital pH / MV Meter

How to Calibrate the IC860031 Benchtop Digital pH / MV Meter

Calibrating your IC860031 digital ph meter regularly and before taking measurements will ensure your device is working correctly. The IC860031 offers two calibration options:

  1. Automatic buffer recognition using National Institute of Standards and Technology (NIST) standard buffers; and
  2. Calibration using custom buffers.

The easy-to-use automatic custom buffers group is the most popular option for Australia because some of the NIST calibration points are not used in this country. For this reason, this guide details how you perform a two-point calibration on the IC860031 pH meter using custom buffers.

Figure 1 Watch an Instrument Choice Scientist calibrating the IC860031 digital pH / MV meter here.

How to Calibrate the IC860031 Benchtop Digital pH / MV Meter 

Tools Required

This article explains how to calibrate the IC860031 digital pH meter to two points, although it can be calibrated to up to five custom points depending on your required accuracy. You’ll need the following gear:

  1. IC860031 Benchtop Digital pH / MV Meter 
  2. Distilled or deionised water (we used Deionised Water)
  3. Two calibration solutions (we used 01 and PH7); and
  4. Three clean containers. 

Custom buffer ranges for the IC860031 include the following:

  • pH 1.00 to 3.00
  • pH 3.50 to 5.50
  • pH 6.00 to 8.00
  • pH 8.50 to 10.50; and
  • pH 11.50 to 13.50

 

How to Perform a Two-Point Calibration

Once your meter is switched on and set up, you can start your calibration. If you are unsure how to assemble your meter, click here.

How to Select the Custom Buffer Range

To adjust to the custom buffer setting, press and hold the SET key for two seconds.

  1. Press the UP arrow until “buF” appears on the top line of the LCD.
  2. Press the ENTER key.
  3. Either “CUSt” or “NIST” will blink on the LCD.
  4. Press the UP arrow key until “CUSt” blinks on the LCD
  5. Press ENTER to confirm your selection.
  6. The device will return to the main settings menu. Press ESC to return to measurement mode.

How to Perform a Two-Point Calibration

After setting the device to custom buffer mode, it’s time to take a measurement.

  1. Pour pH7, pH4 and deionised water into each of the three clean containers. Ensure it is enough to submerge the electrode.
  2. Rinse the electrode in deionised water, then shake gently to dry.
  3. Immerse the electrode into pH7 buffer solution, stirring the probe gently to ensure a uniform sample.
  4. Press CAL to enter into calibration mode. On the LCD, you will see:
    1. The measured value
    2. The CUSTOM buffer value; and
    3. CAL, which indicates you are now in pH calibration mode.
  5. You will need to adjust the CUSTOM value (B) to match the calibration buffer standard in use – i.e., 7.00 for this example. To do this, press HOLD to move through the available buffer ranges. The screen displays the midpoint of each range (see table 1 below for an example). So for pH 7 buffer solution, press HOLD until 7.00 appears on the second line.

    Custom Buffer Ranges

    pH 1.00 to 3.00

    pH 3.50 to 5.50

    pH 6.00 to 8.00

    pH 8.50 to 10.50

    pH 11.50 to 13.50

    Displayed midpoint

    2.00

    4.50

    7.00

    9.50

    12.50

    Table 1 The displayed midpoint for each custom calibration buffer range

  6. If the reading is stable (indicated by READY displayed in the top left corner), press ENTER to calibrate to pH7.00.
  7. The screen will blink as calibration is occurring.
  8. Once complete, the device will return to calibration mode. This time the measured value will be the same as the CUSTOM buffer value.
  9. Remove the probe from the buffer solution and rinse in the deionised water.
  10. Next, immerse the electrode into the second buffer solution, pH 4.01.
  11. Press HOLD until the closest range to the pH buffer solution appears (refer to table 1) – in this case, pH 4.50.
  12. Then adjust the CUSTOM BUFFER using the UP and DOWN arrows until the custom buffer reads 4.01, which matches the solution.
  13. Press enter to confirm and begin the calibration to this buffer point.
  14. As with the first calibration point, both lines on the screen will blink as the unit calibrates.
  15. Once complete, the device will return to calibration mode. This time the measured value will be the same as the buffer value (or within the expected range ±02).
  16. Two-point calibration is complete. From here, you can continue and calibrate the unit up to a total of five points. Or exit calibration mode by pressing the CAL/ESC button.
  17. The device will return to normal pH measurement mode and is ready to take accurate pH measurements.

Conclusion

The ability to calibrate to custom buffers on the IC860031 digital pH meter ensures you are not limited in buffer choices and can select the most suitable solutions for your application.

Need a hand with your IC860031 Benchtop Digital pH/MV meter? See the links below or contact us!

Additional links

See our other newsletter articles here!

Contact our expert scientists now to get the right meter or data logger to suit your needs and discuss your project. 

Instrument Choice will respond in 3 business hours

Get in touch with our expert scientists Instrument Choice.

In less than 3 business hours you will have an answer from our expert scientists.

Our Australian experts have access to a huge range of products from national and international suppliers.

If for any reason you have trouble using this form please email the following address and your enquiry will still receive a response from a scientist within 3 business hours: scientists@instrumentchoice.com.au